Precision and agreement of axial length in paediatric population measured with MYAH and AL-Scan biometers

AUTHORS:

Aldaba, M., Ochando, P., Vila-Vidal, N., Vinuela-Navarro, V., Guisasola, L., & Perez-Corral, J. 

PUBLICATION:

Clin Exp Optom. 2024 Sep;107(7):748-753. doi: 10.1080/08164622.2023.2277287. Epub 2023 Nov 13.

Study Purpose

To assess the precision (repeatability and reproducibility) and agreement of the MYAH and AL-Scan biometers in a paediatric population

Overview

Study Design

    Prospective study

# of Eyes/
Patients

Right eyes of 187 subjects; mean age was 8.5 ± 0.3 years, ranging from 7.9 to 9.8 years

Study Device

  • MYAH (Topcon, Japan)
  • AL-Scan (Nidek, Japan)

Outcome Measures

Repeatability and reproducibility of mean axial length measurements and agreement of the 2 devices

Results

  • For the repeatability study, the within-subject standard deviation was 0.01 mm, and the repeatability limit was 0.04 mm for both instruments, with no statistically significant differences among repeated measures (p = 0.162 for MYAH and p = 0.774 for AL-Scan)
  • For the reproducibility study, the within-subject standard deviation was 0.01 mm and the repeatability limit was 0.04 mm. There were statistically significant differences for the repeated measures for the AL Scan (p = 0.002) but not for the MYAH (p = 0.643)
  • For the agreement study between both instruments, the 95% limit of agreement ranged from −0.04 to 0.05 mm, and the differences were statistically significant (p = 0.021). The comparison has a LoA of approximately ±0.04 mm, and thus the difference among instruments was less than this value in 95% of the cases

IMOvifa reduced measurement time by

39%

FIGURE 1

FIGURE 1.

Bland and Altman plot for the agreement study between MYAH and AL-Scan. The difference between measurements (MYAH minus AL-Scan) is plotted against the mean value of both measurements. The dash-dotted line indicates the mean difference and the dashed lines the upper and lower LoA. Adapted from Clin Exp Optom. 2024 Sep;107(7):748- 753. doi: 10.1080/08164622.2023.2277287. Epub 2023 Nov 13.

Conclusions

• In the analysis of the precision of the MYAH and AL-Scan in paediatric population, the repeatability and reproducibility have been found to be equal • The good agreement between instruments suggests that the two devices can be used interchangeably for myopia control

Product Comparison

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Corneal Topography

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Size and Weight

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Lorem Ipsum

X.X”

360° Rotating
Touchscreen

Control Lever

X.X”

LCD Screen Panel

Wavefront

Control Lever

8.5″

LCD Screen Panel

Lorem Ipsum

EMR

EMR Connectivity

Corneal Aberrometry

Ocular Aberrometry

Pupillometry

Tear Film Assessment

Viewer Software Available

EMR

EMR Connectivity

EMR

EMR Connectivity

The Topcon Healthcare Ophthalmology Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

The Topcon Healthcare Myopia Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.