Comparison of the TEMPO Binocular Perimeter and Humphrey Field Analyzer

AUTHORS:

Nishida T, Weinreb R, Arias J, Vasile C, Moghimi S.

PUBLICATION:

Sci Rep. 2023 Dec 1;13(1):21189.

Study Purpose

To compare the TEMPO 24–2 Ambient Interactive Zippy Estimated by Sequential Testing (AIZE)-Rapid with the Humphrey field analyzer (HFA) 24–2 Swedish Interactive Threshold Algorithm (SITA)-Fast

Overview

Study Design

    Prospective

# of Eyes/
Patients

740 eyes of 370 participants, including 68 healthy, 262 glaucoma suspects, and 410 glaucoma

Study Device

  • TEMPO binocular perimeter
  • Humphrey® field analyzer (HFA)
  • Cirrus® optical coherence tomography (OCT)

Outcome Measures

Visual field (VF) parameters, reliability indices, and coefficient of determinations for VF parameters and retinal nerve fiber layer (RNFL) thickness; measurement time and participant survey

Results

  • No significant differences were seen in mean deviation and visual field index between the two perimeters (P > 0.05), while significant differences were seen in pattern standard deviation and foveal threshold between TEMPO and HFA (both P < 0.001)
  • TEMPO AIZE-Rapid demonstrated lower values than HFA SITA-Fast for all reliability indices: fixation loss (11.2 [10.0, 12.5] % for HFA and 8.9 [7.3, 10.5] % for TEMPO), false positive (4.1 [3.7, 4.6] % for HFA and 1.3 [1.1, 1.5] % for TEMPO), and false negative (4.3 [3.8, 4.8] % for HFA and 0.4 [0.3, 0.4] % for TEMPO)
  • A stronger association between VF mean sensitivity (dB or 1/L) and
  • circumpapillary RNFL was found for TEMPO (adjusted R2 = 0.25; Akaike information criteria [AIC] = 5235.5 for dB, and adjusted R2 = 0.29; AIC = 5200.8 for 1/L, respectively) compared to HFA (adjusted R2 = 0.22; AIC = 5263.9 for dB, and adjusted R2 = 0.22; AIC = 5262.7 for 1/L, respectively) (Figure 1)
  • Stronger associations were found in the inferotemporal disc, followed by
  • modest associations in the superotemporal disc area. Weak associations were observed in the temporal disc area
  • Measurement time was faster for TEMPO compared to HFA
  • (261 s vs. 429 s, P < 0.001)
  • 73% of participants preferred TEMPO while 17% preferred HFA. 83%
  • of participants reported no difficulties with TEMPO; TEMPO received positive feedback in terms of screen readability, ease of concentration, and shorter test duration, as compared to HFA

IMOvifa reduced measurement time by

39%

FIGURE 1

FIGURE 1.

Scatterplots showing associations between global VF mean sensitivity from TEMPO and HFA in dB scale (A, B) and unlogged 1/L scale (C, D), and circumpapillary retinal nerve fiber layer thickness.

Conclusions

TEMPO demonstrated a stronger structure–function association with Cirrus OCT compared to HFA, both globally and sectorally.

Product Comparison

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Corneal Topography

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Size and Weight

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Lorem Ipsum

X.X”

360° Rotating
Touchscreen

Control Lever

X.X”

LCD Screen Panel

Wavefront

Control Lever

8.5″

LCD Screen Panel

Lorem Ipsum

EMR

EMR Connectivity

Corneal Aberrometry

Ocular Aberrometry

Pupillometry

Tear Film Assessment

Viewer Software Available

EMR

EMR Connectivity

EMR

EMR Connectivity

The Topcon Healthcare Ophthalmology Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

The Topcon Healthcare Myopia Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.