Robotic Spectral-domain versus Swept-source OCT in the association between OCT-Angiography metrics and glaucoma severity

AUTHORS:

Fazio MA, Clark ME, Girkin CA

PUBLICATION:

Invest. Ophthalmol. Vis. Sci. 2024;65(7):1212.

Study Purpose

To assess the association between OCT-Angiography (OCTA) metrics and glaucoma severity in subjects imaged by a Spectral-domain (SD) robotic OCT and a swept-source (SS) OCT

Overview

Study Design

    Prospective, observational

# of Eyes/
Patients

Thirty-four eyes of 26 glaucoma subjects

Study Device

    Maestro2 SD-OCT (Topcon Healthcare, Tokyo, Japan) Triton SS-OCT (Topcon Healthcare, Tokyo, Japan) *investigational OCTA acquisition software

Outcome Measures

Linear correlation between radial peripapillary capillary vessel density and visual field mean defect (MD)

Results

  • Radial peripapillary capillary density showed significant association with VF MD for both robotic SD-OCT (2.067 dB/%, p=0.001655) and SS-OCT (1.94 dB/%, p=0.004026); both associations were independent of layer thickness
  • Strength of the correlation was higher for the SS-OCT (R2=0.54) than the SD-OCT (R2=0.45) (Figure 1)

IMOvifa reduced measurement time by

39%

FIGURE 1

FIGURE 1.

Linear correlation between radial peripapillary capillary vessel density and visual field mean defect (MD). Scatterplots are color-coded by scan quality’s native score (Q-mean). Adapted from Invest. Ophthalmol. Vis. Sci. 2024;65(7):1212

Conclusions

• Investigational OCTA metrics by Maestro2 (robotic SD-OCT) performed comparably with the Triton’s (SS-OCT) OCTA metrics in their association with glaucoma severity • The highly automated capabilities of the Maestro2’s OCTA may make it a valuable tool for primary eye care and population screening programs

Product Comparison

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Corneal Topography

Auto Measure

Auto Right/Left

Full Auto
Align/Measurement

Rotary Prism
Autorefraction

Keratometry

Auto Measure

Size and Weight

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Height
XX Inches

Width
XX Inches

Depth
XX Inches

Weight
XX Pounds

Lorem Ipsum

X.X”

360° Rotating
Touchscreen

Control Lever

X.X”

LCD Screen Panel

Wavefront

Control Lever

8.5″

LCD Screen Panel

Lorem Ipsum

EMR

EMR Connectivity

Corneal Aberrometry

Ocular Aberrometry

Pupillometry

Tear Film Assessment

Viewer Software Available

EMR

EMR Connectivity

EMR

EMR Connectivity

The Topcon Healthcare Ophthalmology Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

The Topcon Healthcare Myopia Solution

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.

Fundus Cameras

Lorem ipsum dolor sit, consectetuer adipiscing elit, sed diam nonummy.